高温凝胶渗透色谱仪中气泡的排除

Trouble-shooting of Bubbles in High Temperature Gel Permeation Chromatography

  • 摘要: 叙述了高温凝胶(渗透)色谱(GPC)的流动相中出现气泡时的现象,指出气泡的出现对仪器运行和样品测试的影响,以及排除气泡的方法和过程.对可能出现气泡的原因进行了分析,讨论了避免出现气泡现象需要注意的问题和采取的措施.

     

    Abstract: Appearance of bubbles in PL-220 high temperature gel permeation chromatography was described. Its effects on the instrument and experimentation were discussed. With these bubbles, the instrument running state became unstable and baseline took on serrated fluctuation. Many reasons could induce the bubbles, such as pipeline leaking, valves slop, and manipulation error. These reasons were analyzed one by one. It was decompose of solvent which resulted in the appearance of bubbles. Many methods could be adopted to avoid or resolve this problem. The instrument should keep idle on small flow rate and lower temperature such as 40℃, a temperature higher than room temperature no more than 20℃. Replacing the circumfluence solvent with antioxidant periodically and keep it fresh.

     

/

返回文章
返回