Study on Locating Sample in Vibrating Specimen Magnetometer Measurement by Physical Property Measurement System

  • 摘要: 采用美国Quantum Design公司生产的综合物性测量系统振动样品磁强计(VSM)测量组件进行磁测量时,由于测试样品的原因,经常会出现MultiVU系统自动反馈位置参数不正确的情况. 通过对其自动反馈的过程进行分析,发现确定的非最佳位置获得的样品反馈磁信号和最佳位置的真实值存在对应的百分比值. 利用高场下强磁信号容易找到位置参数的特性,举例说明不用增加背景磁场,对自动反馈位置参数不正确的样品进行正常测试并推算百分比值,再利用该百分比值进行数据修正的方法. 方法确保测试数据的可靠性,减少不必要的时间浪费,可以提高测试效率.

     

    Abstract: When the vibrating specimen magnetometer (VSM) user kit of physical property measurement system (PPMS) produced by Quantum Design is used for the magnetic measurement, it is usually shown that the MultiVU software cannot feedback the correct position parameter due to different kinds of test samples. Through the analysis of the automatic feedback the process, it was found that there was a percentage value corresponding to the feedback magnetic signal of the sample obtained from the determined non-optimal position and the real value of the optimal position. According to the convenience of getting the position parameter by the strong magnetic signal in high field, examples were given to illustrate a method for the data correction without adding the background field, which was performing normal testing and extrapolating percentage values for samples with incorrect automatic feedback position parameters, and then using the percentage values for data correction. The method can ensure the reliability of measurement data, reduce unnecessary waste of time, and improve the test efficiency.

     

/